RxPG News Feed for RxPG News

Medical Research Health Special Topics World
  Home
 
   Health
 Aging
 Asian Health
 Events
 Fitness
 Food & Nutrition
 Happiness
 Men's Health
 Mental Health
 Occupational Health
 Parenting
 Public Health
 Sleep Hygiene
 Women's Health
 
   Healthcare
 Africa
 Australia
 Canada Healthcare
 China Healthcare
 India Healthcare
 New Zealand
 South Africa
 UK
 USA
 World Healthcare
 
   Latest Research
 Aging
 Alternative Medicine
 Anaethesia
 Biochemistry
 Biotechnology
 Cancer
 Cardiology
  CAD
  CHF
  Clinical Trials
  Hypertension
  Myocardial Infarction
 Clinical Trials
 Cytology
 Dental
 Dermatology
 Embryology
 Endocrinology
 ENT
 Environment
 Epidemiology
 Gastroenterology
 Genetics
 Gynaecology
 Haematology
 Immunology
 Infectious Diseases
 Medicine
 Metabolism
 Microbiology
 Musculoskeletal
 Nephrology
 Neurosciences
 Obstetrics
 Ophthalmology
 Orthopedics
 Paediatrics
 Pathology
 Pharmacology
 Physiology
 Physiotherapy
 Psychiatry
 Radiology
 Rheumatology
 Sports Medicine
 Surgery
 Toxicology
 Urology
 
   Medical News
 Awards & Prizes
 Epidemics
 Launch
 Opinion
 Professionals
 
   Special Topics
 Ethics
 Euthanasia
 Evolution
 Feature
 Odd Medical News
 Climate

Last Updated: Oct 11, 2012 - 10:22:56 PM
Cardiology Channel

subscribe to Cardiology newsletter
Latest Research : Cardiology

   EMAIL   |   PRINT
Researchers did not detect interference from PDAs with pacemakers

Dec 23, 2004 - 3:15:00 PM

 
[RxPG] With the dynamic evolution of wireless technology, Mayo Clinic researchers have been concerned about the potential effects of electromagnetic interference on heart pacemakers and implantable cardioverter-defibrillators. In the current issue of Mayo Clinic Proceedings, researchers report they did not detect interference from personal digital assistants (PDAs).

The findings are important because wireless communication has grown and advanced quickly. Hospitals and clinics have installed wireless local area networks (WLAN), which enable users to establish a wireless network connection with computers or other data devices throughout a building or multiple buildings that have the necessary data infrastructure in place. The wireless capabilities allow physicians and other health care professionals immediate access to a variety of information when evaluating and treating patients. Patients also are carrying wireless devices and need to understand if there would be any adverse reactions to implantable cardiac devices.

David Hayes, M.D., a Mayo Clinic physician and lead researcher in the study, says researchers did not expect to find interference based on their past experiences with other devices they have tested.

"When new devices are used near a patient with a life-sustaining implantable device, there is a potential of electromagnetic interference, and assessment of potential interactions is critical," says Dr. Hayes. "Despite the increasing sophistication of sensing circuitry in contemporary pacemakers and implantable cardioverter-defibrillators, these devices are still susceptible to electromagnetic interference and physicians need good data telling them which ones are or aren't. And as technology advances, we'll need continual testing to stay up to date."

In the Mayo Clinic study, testing was conducted between March 6 and July 30, 2003, using devices outside of the body. The cardiac devices were exposed to an HP Compaq iPAQ Pocket PC personal digital assistant fitted with a Cisco Aironet WLAN card. The testing of devices that had not been implanted was done for clinical safety purposes, says Dr. Hayes. However, he says testing of devices inside a patient's body would be easy to design and conduct.

"Such testing is necessary to provide definitive answers for individual patients," says Dr. Hayes. "For example, a patient who is pacemaker dependent may ask whether a specific WLAN device can be used and/or carried safely in a coat pocket when turned on near the patient's implanted device."

Dr. Hayes suggests that a template for further study could be from the cellular telephone study he led which was published in 1997 in the New England Journal of Medicine. That study tested numerous implanted devices with the most commonly available cellular telephones and with cellular telephones not yet commercially available but representing a different "frequency" or design. Testing for the PDA study followed methods provided in the American National Standards Institute/Association for the Advancement of Medical Instrumentation Pacemaker Committee protocol.

"As other wide local area network products are developed and made commercially available, additional testing will be needed to ensure clinical safety," Dr. Hayes says.

Researchers involved in the study were Jeffrey Tri, Jane Trusty and Dr. Hayes.



Publication: A peer-review journal, Mayo Clinic Proceedings

Advertise in this space for $10 per month. Contact us today.


Related Cardiology News
New NIH-funded resource focuses on use of genomic variants in medical care
World Heart Day 2013
The higher the better?
Common blood pressure drug reduces aortic enlargement in Marfan syndrome
Cardiovascular risk factors highest in winter and lowest in summer
Quitting smoking drops heart attack risk to levels of never smokers
Study finds mechanical chest compressions are equally as effective as manual CPR
Impact of AF on stroke risk eliminated with multiple risk factors
Mass screening identifies untreated AF in 5% of 75-76 year olds
Diabetic stroke risk after AMI drops in 10 year period

Subscribe to Cardiology Newsletter

Enter your email address:


 Additional information about the news article
A peer-review journal, Mayo Clinic Proceedings publishes original articles and reviews dealing with clinical and laboratory medicine, clinical research, basic science research and clinical epidemiology. Mayo Clinic Proceedings is published monthly by the Mayo Foundation for Medical Education and Research as part of its commitment to the medical education of physicians. The journal has been published for more than 75 years and has a circulation of 130,000 nationally and internationally. Copies of the articles are available online at www.mayoclinicproceedings.com.
 Feedback
For any corrections of factual information, to contact the editors or to send any medical news or health news press releases, use feedback form